7.2 Electrical Mounting
7.2.1 Cable
Solder a shielded cable with diameter 7-8mm and a core cross-section of 0.5...0.75mm² to
the enclosed coupling socket.
•
The number of necessary wires is given in the connection diagram (chapter 7.3).
7.2.1.1 Cable Recommendation
No. of wires/ diameter / type / cable diameter
CABLE 8x0.5mm² LIYCY BLACK, UV- resistant, Ø 7.6mm
7.2.2 Cable Shield
The connection of the cable shield between sensor and data acquisition device should be se-
lected in way, that in case of over-voltages no equalizing currents will flow that might destroy
the electronic components.
The connection of the cable shield should depend on the selected isolated, or respectively,
non-isolated mounting of the sensors.