SINGLE-BAY SMART CHARGER

INSTRUCTIONS FOR USE

ENGLISH (EN) – ITALIANO (IT)

60501

This

Instructions For Use manual

is the most comprehensive source of information

for the safe and effective use of your product. Keep and consult this reference

manual during the life of the product.

IMPORTANT INFORMATION

The words WARNING, CAUTION, and NOTE have special meaning and should be 

reviewed.

COMPLIANCE STATEMENT

Medical Device acc. Directive 93/42/EEC

This device is subject to the conformity assessment procedure according to annex 

VII of Directive 

93/42/EEC.

INDICATION FOR USE

This smart charger is a component of the

ViVi®

System, to be used by healthcare

professionals outside of the patient environment to charge the Battery Pack HW

REF 60402.

CONTRAINDICATIONS

None known. 

FOR USE WITH

Use this smart charger to charge the Battery Pack HW REF 60402.

USER/ PATIENT SAFETY

WARNINGS: 

This battery charger may not be located in the vicinity of a patient. In

accordance with requirements defined in EN60601-1-1, the horizontal

distance to a patient must be at least 1.5 m, the vertical at least 2.5 m.

Use the battery charger only in dry indoor environments! Wet and moisture

covered products should not be used!

FCC Advice, valid for USA:

This product has been tested and found to comply

with the limits for a Class B digital device, pursuant to part 15 of the FCC

Rules. These limits are designed to provide reasonable protection against

harmful interference in a residential installation. This equipment generates,

uses and can radiate radio frequency energy and, if not installed and used in

accordance with the instructions, may cause harmful interference to radio

communications. However, there is no guarantee that interference will not

occur in a particular installation.

DO NOT sterilize.

MAKE SURE that the power supply is easily accessible when connected to the

mains socket.

© THI GmbH 2020/06

REF 20605 / Rev.01

WARNING

WARNING

highlights a safety-related issue. ALWAYS comply with 

this information to prevent patient and/or healthcare staff injury. 

CAUTION

CAUTION

highlights a product reliability issue. ALWAYS comply 

with this information to prevent product damage.

NOTE

NOTE

supplements and/or clarifies procedural information.

SYMBOL DEFINITIONS

The symbols located on the equipment and/or labelling are defined in this

section.

HOW TO USE 

MAKING CONNECTIONS

1. Connect the cable from the power supply to the battery charger.

2. Connect the power cable to the power supply and plug it into the mains socket.

3. The charger does perform as self test. After completion, the LED is going out.

The battery charger is ready for use.

4. If a red LED remains on and no battery is in the charging bay, the battery

charger is defective.

CHARGING

1. Connect the battery charger as described above.

2. Place a discharged Battery Pack into the battery charger.

3. The battery charger runs a battery recognition and initialization cycle. LED’s

might go on and off during this cycle.

4. The battery will be charged. Leave the battery until the solid green LED

appears.

NOTE:

We recommend leaving the battery pack in the charging bay if it is

intended for use.

CLEANING

1. Unplug the power supply from the mains socket.

2. Wipe the battery charger and the power supply with a soft cloth dampened 

with a non-abrasive hospital disinfectant. Follow applicable hospital protocol for 

cleaning.

3. Dry with a lint-free towel or medical-grade compressed air.    

THI

- Total Healthcare Innovation GmbH Gewerbestraße 4

9181 Feistritz im Rosental AUSTRIA

phone: +43 4228 30100

fax: +43 4228 30100-20

mail: offi[email protected]

web: www.thigmbh.at

SYMBOL

DEFINITION

Caution, consult accompanying documents
Consult instructions for use
Mark of conformity to applicable European Directives
Manufacturer
Batch number

QTY

Quantity 
Article number
Non sterile
Use only in enclosed spaces (indoors)
Dispose acc. to WEEE directive (EU) or acc. local guidelines 
DC Voltage
China RoHs
Compliance to California Energy Commission requirements

INSTRUCTIONS FOR USE - ENGLISH

page 1 of 2

SPECIFICATION

Power Supply

Single-Bay Smart Charger

REF (Type)

30601

60501

Input Voltage

100-240VAC/ 50-60 Hz

19-26VDC / 3.4 A max

Charging and Output 

Voltages

19VDC ±5% / 3.4A

0 – 17.4 VDC

0 – 4.8 A

Power

65W

50W

Dimensions (LxWxH)

95 x 51 x 25.4 mm

151 x 90 x 43 mm

Operating / Storage 

Temperature

0°C to 40°C / 

-10°C to 70°C

0°C to 40°C / 

-10°C to 70°C

Weight

270g

198g

Safety/EMC Standards

*..valid only for power supply

EN60950-1 / UL60950-1* / EN55011 / EN55022 / 

EN55024* / FCC15 Class B/ EN61000-4-2,-3,-4,-5,-6,-8,-

11* / EN61000-3-2,3-3*, EN60601-1 / EN60601-1-2

CHARGING INDICATION (LED)
One time Red/ 

Orange/ Green

Self test: Charger is ready for use (after LED goes out)

Red/ Green 

blinking

The battery is not recognized as Smart Battery. Either a

conventional battery has been inserted or an extremely

discharged Smart Battery. If it is a Smart Battery, it will be

reactivated within 15 minutes and recharged. If it is not the

case, the LED with turn to solid red – see below.

Orange blinking

The battery is currently being calibrated.

Solid Orange

The inserted battery is the correct type and is currently being

charged.

Solid Green

The battery is fully charged and can be removed for use.

Red blinking

The battery is too hot or too cold to be charged without

damage. If the battery is too cold it will be charged as soon as it

is warmed up sufficiently. If the battery is too hot it should be

removed to cool down.

Solid Red

Either:

The battery is badly damaged and needs to be replaced or

it is a conventional battery which can not be charged