background image

26-1-16 

EXPLANATION OF CONTROLS 
 

The controls are located in a control enclosure to the right hand side of the bath. 
 
 

MAIN SWITCH: 

 

This is a combination power switch and circuit breaker, when pressed this switch 
allows power to be supplied to the controls. Incorporated in the face of the switch is a 
“MAINS ON” light. 
 

CAUTION: 

The equipment is fitted with a combination 

circuit breaker/switch. If the switch continues to trip, it 
may indicate an electrical fault exists. Switch off the 
power on consult a qualified technician. 

 
 
TEMPERATURE CONTROL/SENSOR: 
 

This is a Brainchild BTC901 analogue control scaled a 2ºC increments and has been 
configured to provide stable control with minimum of overshoot and quick recovery 
time.  
 
Sensor type is a RTD Pt100. 
 

Maximum recommended temperature is 60ºC. 
 
 

 

 

HEAT CYCLE INDICATOR: 

 
This output light is illuminated when power is applied to the heating element.  
Green: Element off.  Red: Element on. 
 

LIGHT SWITCH: 

 
This is a push button, press to turn light on, press again to turn light off. When 
activated, the fluorescent light will be illuminated. 
 
 
 

Содержание TFB-1

Страница 1: ...fic Equipment Pty Ltd ABN 80 000 859 129 10 12 Ross Place Wetherill Park NSW 2164 P O Box 6862 Wetherill Park DC NSW 1851 Phone 02 9604 3911 International 61 2 9604 3911 Fax 02 9725 1706 International...

Страница 2: ...rts apply when there is a Possibility of personal injury Caution alerts apply when there is a Possibility of damage to the Equipment To maintain safe operation of this equipment be aware of the ALERT...

Страница 3: ...king material check that it is complete Assemble the components and inspect for damage Retain the packaging materials until the equipment has been thoroughly tested Notify the detail of any damage wit...

Страница 4: ...temperature is set at the temperature controller The temperature sensor is located within a spring clip and is designed to rest against the glass insert dish The signal from the temperature sensor is...

Страница 5: ...y be removed for cleaning The external case and the stainless steel surfaces may be wiped clean using a damp soft cloth Note Stainless steel is under most conditions extremely resistant to corrosion T...

Страница 6: ...icate an electrical fault exists Switch off the power on consult a qualified technician TEMPERATURE CONTROL SENSOR This is a Brainchild BTC901 analogue control scaled a 2 C increments and has been con...

Страница 7: ...to the well and fill it with water to a maximum depth of 40mm CAUTION Ensure that the ventilation for the rear cooling fan are not obstructed 2 Place the main switch in the ON position reset and immed...

Страница 8: ...ng the preceding clause and to the extent permissible by law the liability of Thermoline is limited in relation to the warranted product and at the option of Thermoline to replacing the product or the...

Отзывы: