Operating Principle
The detector is designed for ultraviolet (UV) and visible (VIS) absorption
spectroscopy in combination with HPLC or UHPLC separations. After
sufficient separation from other sample compounds, the analysis of the
target compound follows the Beer-Lambert law. This means that the
response of the detector is proportional to the concentration of the
analyte.
The following picture shows the optics of the detector, and illustrates
how the detector operates:
Figure 1: Operating principle of the optics
No. Description
1
Deuterium lamp
2
Lamp mirror
3
Filter paddle (shutter)
4
Flow cell
5
Spectrograph mirror
6
Entrance slit
7
Grating
8
Diode array
Содержание Vanquish VH-D10
Страница 4: ...Contacting Us Page 4 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 44: ...4 Unpacking Page 44 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 80: ...5 Installation Page 80 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 110: ...6 Operation Page 110 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 162: ...7 Maintenance and Service Page 162 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 190: ...11 Appendix Page 190 Vanquish Diode Array Detectors VH D10 Operating Manual...
Страница 191: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 191 12 Index...
Страница 195: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 195...
Страница 196: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...