9
•
Specifications
Page 174
Diode Array/Multiple-Wavelength Detectors (VC-D11, VC-D12, VF-D11)
Operating Manual
9.2
Physical Specifications
The physical conditions of the device are specified as follows:
Type
Specification
Range of use
Indoor use only
Ambient operating
temperature
5 °C - 35 °C
Ambient storage temperature
-20 °C - 45 °C
Ambient operating humidity
20% - 80% relative humidity (non-condensing)
Ambient storage humidity
Maximum 60% relative humidity (non-condensing)
Operating altitude
Maximum 2000 m above sea level
Pollution degree
2
Power requirements
100 – 240 V AC, ± 10 %; 50/60 Hz; max. 245 W / 255 VA
Overvoltage category
II
Emission sound pressure level
typically < 55 dB(A)
Dimensions
(height x width x depth)
15.9 x 42 x 62 cm
Weight
Approx. 17 kg
Содержание Vanquish VC-D11
Страница 4: ...Contacting Us Page 4 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 10: ...Contents Page 10 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 42: ...3 Device Overview Page 42 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 48: ...4 Unpacking Page 48 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 168: ...8 Troubleshooting Page 168 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 186: ...11 Appendix Page 186 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 191: ...Index Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual Page 191...
Страница 192: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...