8
•
Troubleshooting
Page 158
Diode Array/Multiple-Wavelength Detectors (VC-D11, VC-D12, VF-D11)
Operating Manual
8.1
General Information about Troubleshooting
The following features help you to identify and eliminate the source for
problems that may arise during operation of the device.
TIP
For information about operating issues that might occur during the
operation of a Vanquish system, refer to the
Vanquish System Operating
Manual
.
If you are unable to resolve a problem following the instructions given
here or if you experience problems that are not covered in this section,
contact Thermo Fisher Scientific Technical Support for assistance. See
the contact information at the beginning of this manual.
To facilitate device identification, have the serial number and technical
name available when communicating with Thermo Fisher Scientific.
Status Indicators
The status indicator LED bar on the front side of the device and the
STATUS
LED on the keypad inside provide quick visual feedback on the
operational status of the device. If the device firmware detects a
problem, the status indicators are red. The problem is reported to the
Chromeleon software and a message appears in the Audit Trail.
Alarms
Leaks are a potential safety issue. Therefore, if a leak sensor detects
leakage, beeping starts to alert you in addition to the message in the
Chromeleon Audit Trail and the status indicators changing to red. Follow
the instructions in this manual to find and eliminate the source for the
leakage.
Chromeleon Audit Trail Messages
If the device firmware detects a problem, the problem is reported to the
Chromeleon software.
The Chromeleon software logs information about all events related to
instrument operation for the current day in an Audit Trail. The Audit
Trail is named with the current date, using the format yyyymmdd. For
example, the Audit Trail for May 15, 2019, is named 20190515.
The Instrument Audit Trails can be found on the ePanel Set (Audit
ePanel). In addition, Audit Trails for each instrument are available in the
Chromeleon 7 Console Data view, in the folder of the instrument.
Содержание Vanquish VC-D11
Страница 4: ...Contacting Us Page 4 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 10: ...Contents Page 10 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 42: ...3 Device Overview Page 42 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 48: ...4 Unpacking Page 48 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 168: ...8 Troubleshooting Page 168 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 186: ...11 Appendix Page 186 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 191: ...Index Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual Page 191...
Страница 192: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...