6
Custom Applications
Measure Custom
70 NanoDrop Eight User Guide
Thermo Scientific
Factor or Extinction coeffi-
cient at 1 cm pathlength
(Formula methods only)
Specify whether to use factor or extinction coefficient to calculate concentra-
tion result:
•
User-defined factor
. Enter
factor
for 1 cm pathlength and use
adjacent drop down list to select appropriate
unit
. Equation below
shows how factor is used to calculate sample concentration:
c = (A * f) / b
where:
c = analyte concentration
A = absorbance in absorbance units (A)
f
= factor (typically 1/
, where
= wavelength-dependent molar
absorptivity coefficient, or extinction coefficient)
b = pathlength in cm (determined at measurement time, then normalized
to 10 mm (1 cm) pathlength equivalent)
•
Extinction coefficient and molecular weight
. Enter
extinction
coefficient
for 1 cm pathlength and use adjacent drop down list to
select appropriate
unit
. Equation below shows how extinction coefficient
is used to calculate sample concentration:
c = A / (
* b)
where:
c = analyte concentration
A = absorbance in absorbance units (A)
= wavelength-dependent molar absorptivity coefficient (or extinction
coefficient)
b = pathlength in cm (determined at measurement time, then normalized
to 10 mm (1 cm) pathlength equivalent)
Notes
:
• Refer to product literature for information about factors and extinction
coefficients for specific materials.
• To set up a method that reports absorbance measurements only, select
Factor or Extinction Coefficient with the factor or extinction coefficient set
to “1”.
• If specified unit for factor or extinction coefficient is based on mass (such
as mg/mL) and specified unit for calculated result is based on molarity
(such as pmol/µL) or vice versa, enter
molecular weight
and use
adjacent drop down list to select appropriate
unit
.
Setting
Available Options
Содержание NanoDrop Eight
Страница 6: ...Contents vi NanoDrop Eight User Guide Thermo Scientific...
Страница 24: ......
Страница 30: ...4 Nucleic Acid Applications 30 NanoDrop Eight User Guide Thermo Scientific This page is intentionally blank...
Страница 40: ...4 Nucleic Acid Applications Measure Oligo DNA or Oligo RNA 40 NanoDrop Eight User Guide Thermo Scientific...
Страница 56: ...5 Protein Applications Measure Protein A280 56 NanoDrop Eight User Guide Thermo Scientific...
Страница 76: ...6 Custom Applications Measure Custom 76 NanoDrop Eight User Guide Thermo Scientific...
Страница 126: ...7 Learning Center Measurement Screen Display Options 126 NanoDrop Eight User Guide Thermo Scientific...
Страница 142: ...8 Maintenance 142 NanoDrop Eight User Guide Thermo Scientific This page is intentionally blank...