4
Nucleic Acid Applications
Measure Oligo DNA or Oligo RNA
Thermo Scientific
NanoDrop Eight User Guide 35
Settings for Oligo DNA and Oligo RNA Measurements
The Oligo setup screen appears after you select the Oligo DNA or Oligo RNA
application from the Nucleic Acids tab on the Home screen. From the Oligo DNA or
RNA measurement screen, select the settings icon
to view
Oligo DNA Setup
or
Oligo RNA Setup
.
Setting
Available Options
Description
Oligo Base
Sequence
for DNA: Use the
G, A, T and C keys to
specify the DNA base
sequence
for RNA: Use the
G, A, U and C keys to
specify the RNA base
sequence
Specify your DNA or RNA base sequence.
click the corresponding keys:
You can also enter base sequence using the keyboard, or by
copy and pasting a sequence from another application.
Each time a base is added to the sequence, the software
calculates the following:
•
Factor
. Constant used to calculate oligonucleotide
concentration in
extinction coefficient and pathlength:
f = 1/(
260
* b)
where:
f
= factor
= molar extinction coefficient at 260 nm in ng-cm/µL
b
=
in cm (1 cm for nucleic acids
measured with the NanoDrop Eight instrument)
Add guanine
Add adenine
Add thymine (DNA)
or uracil (RNA)
Add cytosine
Base sequence
Содержание NanoDrop Eight
Страница 6: ...Contents vi NanoDrop Eight User Guide Thermo Scientific...
Страница 24: ......
Страница 30: ...4 Nucleic Acid Applications 30 NanoDrop Eight User Guide Thermo Scientific This page is intentionally blank...
Страница 40: ...4 Nucleic Acid Applications Measure Oligo DNA or Oligo RNA 40 NanoDrop Eight User Guide Thermo Scientific...
Страница 56: ...5 Protein Applications Measure Protein A280 56 NanoDrop Eight User Guide Thermo Scientific...
Страница 76: ...6 Custom Applications Measure Custom 76 NanoDrop Eight User Guide Thermo Scientific...
Страница 126: ...7 Learning Center Measurement Screen Display Options 126 NanoDrop Eight User Guide Thermo Scientific...
Страница 142: ...8 Maintenance 142 NanoDrop Eight User Guide Thermo Scientific This page is intentionally blank...