![ThermoFisher Scientific NanoDrop Eight Скачать руководство пользователя страница 130](http://html1.mh-extra.com/html/thermofisher-scientific/nanodrop-eight/nanodrop-eight_user-manual_1105612130.webp)
8
Maintenance
Maintaining the Pedestals
130 NanoDrop Eight User Guide
Thermo Scientific
Supplies needed
• lint-free laboratory wipes
• deionized water (DI H
2
O)
• for thorough cleaning:
0.5M HCl
To clean the pedestals between
measurements
Lift the instrument arm and clean the
upper and lower pedestals with a new
laboratory wipe.
To clean the pedestals between
users
1. Lift the arm and clean all pedestals
with a new laboratory wipe.
2. Pipette 3–5 µL DI H
2
O onto the
lower pedestal.
3. Lower the arm and wait 2–3
minutes.
4. Lift the arm and clean all pedestals
with a new wipe.
Tip
: When thorough cleaning is
required (for example, to remove dried
sample left on the pedestals),
substitute 0.5M HCl for the DI H2O in
the procedure above and follow with
3-5 µL DI H2O. You can also
compound.
Содержание NanoDrop Eight
Страница 6: ...Contents vi NanoDrop Eight User Guide Thermo Scientific...
Страница 24: ......
Страница 30: ...4 Nucleic Acid Applications 30 NanoDrop Eight User Guide Thermo Scientific This page is intentionally blank...
Страница 40: ...4 Nucleic Acid Applications Measure Oligo DNA or Oligo RNA 40 NanoDrop Eight User Guide Thermo Scientific...
Страница 56: ...5 Protein Applications Measure Protein A280 56 NanoDrop Eight User Guide Thermo Scientific...
Страница 76: ...6 Custom Applications Measure Custom 76 NanoDrop Eight User Guide Thermo Scientific...
Страница 126: ...7 Learning Center Measurement Screen Display Options 126 NanoDrop Eight User Guide Thermo Scientific...
Страница 142: ...8 Maintenance 142 NanoDrop Eight User Guide Thermo Scientific This page is intentionally blank...