iSort
™
Automated Cell Sorter User Guide
55
Damper Flush
The Damper Flush function removes contamination from the pulse damper in the
sample line.
Required solutions
Sheath fluid (1X PBS, flow cytometry grade)
Run Damper Flush
1.
Select
Damper Flush
from the
Maintenance
screen.
2.
Place a new tube containing 3 mL of sheath fluid in the sample tube holder.
3.
Move the sample tube holder up until it clicks into place, then press
Continue
.
The instrument performs the damper flush automatically, which takes
approximately 30 seconds to complete.
4.
When completed, press
Done
to go to the Maintenance screen.