iSort
™
Automated Cell Sorter User Guide
Debubble
The Debubble function is a user-initiated function for clearing air bubbles from the
fluidics lines of the system. Air bubbles in the fluidics lines may cause fluctuations
in the system pressure, which can result in problems with the stream from the
nozzle and adversely affect droplet formation.
Run Debubble
1.
Ensure that the sample tube holder contains an empty sample tube, then select
Debubble
from the Maintenance screen to initiate the Debubble procedure.
The instrument performs the Debubble procedure automatically, which takes
approximately 2 minutes to complete.
2.
When completed, press
Done
to return to the Maintenance screen.