8
Methods
Electrophoresis of E-PAGE
™
Gels
Guidelines for Sample Preparation
Introduction
For optimal results using the E-PAGE
™
System, follow the guidelines for
preparing your protein samples as described in this section.
We recommend that you read this section before preparing your samples. Use
the sample preparation method and loading buffer appropriate for your
detection method:
Application
Method
Loading Buffer
Routine staining and
western blotting
1 (page 11)
Loading Buffer 1
Lumio Green Detection
2 (page 12)
Lumio
™
Gel Sample Buffer
Materials Needed
The following items are needed for sample preparation. See page 71 for
ordering information.
•
Protein sample
•
NuPAGE
®
Sample Reducing Agent (10X)
•
E-PAGE
™
Loading Buffer 1 (4X) included in the kit
•
Deionized water
•
Heating block set at 70
°
C
•
Molecular weight markers (page 10)
•
(
Optional
)
Lumio
™
Green Detection Kit for detection of Lumio
™
fusion
proteins (page 12)
E-PAGE
™
Gels contain SDS and are designed for performing electrophoresis
under denaturing conditions.
To obtain the best results, we recommend performing SDS-PAGE under
reducing conditions. If you need to perform SDS-PAGE under non-reducing
conditions,
do not
add NuPAGE
®
Sample Reducing Agent (10X) during
sample preparation.
Содержание E-PAGE Gels