DeviceLink Connect
Technical Specifications | 7
Independent Sensor Specifications
Temperature (RTD)
Amperage
(Current Transducer)
Auxiliary Sensor/Door
Switch or Dry Contact
Interface
USB
USB
RS485 (RJ45)
Type
3 Wire, 100 Ohm Platinum
RTD (TC = 3850 ppm/K)
Split core current transducer
with built-in transient voltage
suppression.
Dry Contact / Door Switch
System Accuracy
±0.5°C (0.0% full scale)
±2%
n/a
Operating Range
-200°C to +600°C
0-50 A or 0-80 A
n/a
Display Resolution
0.1
0.1
n/a
System Resolution
0.01
0.01
n/a
Temperature/RH
Temperature/RH
CO
2
Interface
USB
RS485 (RJ45)
RS485 (RJ45)
System
Accuracy
Temp: ±1°C
RH: ±3% RH
Temp from +10°C to +30°C: ±0.5°C
Temp from -40°C to +10°C: ±0.6°C
RH at temps from 0°C to +40°C and
0% to 90% RH: ±3% RH
RH at temps from 0°C to +40°C and
90% to 100% RH: ±5% RH
RH at temps from -40°C to 0°C,
+40°C to 60°C and 0% to 90%
RH: ±5% RH
RH at temps from -40°C to 0°C,
+40°C to 60°C and 90% to 100%
RH: ±7% RH
Temp at 25°C, Pressure at 1013
hPa, and CO
2
at 5% = ±0.1% CO
2
CO
2
at 0% to 8% = ±0.2% CO
2
CO
2
at 8% to 20% = ±0.4% CO
2
Operating
Range
Temp: -17°C to +49°C
RH: 2% to 95% RH
(non-condensing)
Temp: -40°C to +60°C
RH: 0% to 100% RH
0% to 20% CO
2
Display
Resolution
0.1
0.1
0.1
System
Resolution
Temp: 0.01
RH: 0.01
Temp: 0.01
RH: 0.01
0.01