4
Rotor Loading
Improper Loading
Thermo Scientific
75003658 TX-200
4-3
Improper Loading
Maximum Loading
The rotor can run at high speeds. The rotor design has sufficient reserve stability even when spinning at
top speed.
The safety system of the centrifuge requires that you do not overload the rotor.
There are two options available for centrifuging samples whose weight, including adaptor, exceeds the
maximum permissible load:
•
Reduce the fill level.
•
Reduce the speed.
Use the table or the formula:
Actual Load (g/bucket)
RPM
max
275
5500
295
5310
315
5139
335
4983
355
4841
375
4710
395
4589
415
4477
435
4373
455
4276
475
4185
495
4099
515
4019
535
3943
555
3872