background image

 

Rotor Loading

Improper Loading

Thermo Scientific

75003658 TX-200

4-3

Improper Loading

Maximum Loading

The rotor can run at high speeds. The rotor design has sufficient reserve stability even when spinning at 
top speed.

The safety system of the centrifuge requires that you do not overload the rotor.

There are two options available for centrifuging samples whose weight, including adaptor, exceeds the 
maximum permissible load:

Reduce the fill level.

Reduce the speed.
Use the table or the formula:

Actual Load (g/bucket)

RPM 

max

275

5500

295

5310

315

5139

335

4983

355

4841

375

4710

395

4589

415

4477

435

4373

455

4276

475

4185

495

4099

515

4019

535

3943

555

3872

Содержание 75003658

Страница 1: ...Thermo Scientific TX 200 Instruction Manual 50121026 d 05 2020 Visit us online to register your warranty thermofisher com labwarranty...

Страница 2: ...WEEE Conformity This product is subject to the regulations of the EU Waste Electrical Electronic Equipment WEEE Directive 2012 19 EU It is marked by the following symbol...

Страница 3: ......

Страница 4: ......

Страница 5: ...5 1 7 120 V 60 Hz Round Bucket 75003655 1 8 Chapter 2 Accessories 2 1 Chapter 3 AutoLock 3 1 Rotor Installation 3 2 Removing the Rotor 3 3 Chapter 4 Rotor Loading 4 1 Before a Run 4 2 Proper Loading 4...

Страница 6: ...Contents ii 75003658 TX 200 Thermo Scientific RCF Values A 1 Chemical Compatibility Chart B 1...

Страница 7: ...manual will result in the expiration of the seller s warranty Scope of Supply If any parts are missing please contact the nearest Thermo Fisher Scientific representative Precautions In order to ensure...

Страница 8: ...articular attention to the following aspects Rotor installation Check that the rotor is locked properly into place before operating the centrifuge Always balance the samples Maximum sample density at...

Страница 9: ...Thermo Scientific 75003658 TX 200 1 1 1 Rotor Specifications Contents Technical Data on page 1 2...

Страница 10: ...me s 30 35 30 35 Sample heating at nmax C referred to ambient temperature of 23 C running time 60 minutes 10 10 Aerosol tight Tested by HPA Porton down UK yes yes Permissible temperature range autocla...

Страница 11: ...40 35 40 Sample heating at nmax C referred to ambient temperature of 23 C running time 60 minutes 9 9 Aerosol tight yes yes Permissible temperature range autoclavable C 121 121 Tested by HPA Porton do...

Страница 12: ...35 30 35 Sample heating at nmax C referred to ambient temperature of 23 C running time 60 minutes 8 8 Aerosol tight Tested by HPA Porton down UK yes yes Permissible temperature range autoclavable C 1...

Страница 13: ...5 40 35 40 Sample heating at nmax C referred to ambient temperature of 23 C running time 60 minutes 9 9 Aerosol tight Tested by HPA Porton down UK yes yes Permissible temperature range autoclavable C...

Страница 14: ...35 30 35 Sample heating at nmax C referred to ambient temperature of 23 C running time 60 minutes 9 9 Aerosol tight Tested by HPA Porton down UK yes yes Permissible temperature range autoclavable C 1...

Страница 15: ...raking time s 30 35 Sample heating at nmax C referred to ambient temperature of 23 C running time 60 minutes 8 Aerosol tight Tested by HPA Porton down UK yes Permissible temperature range autoclavable...

Страница 16: ...king time s 35 40 Sample heating at nmax C referred to ambient temperature of 23 C running time 60 minutes 9 Aerosol tight Tested by HPA Porton down UK yes Permissible temperature range autoclavable C...

Страница 17: ...Thermo Scientific 75003658 TX 200 2 1 2 Accessories Contents Rotor data on page 2 2 Accessories on page 2 2...

Страница 18: ...Oak Ridge Tube 10 5500 32 x 126 Included 10 PP Sealing 75003815 4 1 45500 30 30 25 KIMAX Glass Tube 6 25 x 118 75003805 4 2 25 Universal Tube 25 5 x 118 75003804 4 1 25 Round Bottom Tube 25 x 118 750...

Страница 19: ...Thermo Scientific 75003658 TX 200 3 1 3 AutoLock Contents Rotor Installation on page 3 2 Removing the Rotor on page 3 3...

Страница 20: ...spindle The rotor clicks automatically into place 3 Check if the rotor is properly installed by lifting slightly on the handle If the rotor can be pulled up then it must be reclamped to the centrifuge...

Страница 21: ...s and press against the green AutoLock button At the same time pull the rotor directly upwards with both hands and remove it from the centrifuge spindle Make sure not to jam the rotor while doing this...

Страница 22: ......

Страница 23: ...Thermo Scientific 75003658 TX 200 4 1 4 Rotor Loading Contents Before a Run on page 4 2 Proper Loading on page 4 2 Improper Loading on page 4 3 Maximum Loading on page 4 3 Cycle Counter on page 4 4...

Страница 24: ...ity chart on page B 1 5 Make sure the tubes or bottles do not touch the bucket lid 6 Check the rotor bolts and apply grease 75003786 before using for the first time 7 Check that each bucket or micropl...

Страница 25: ...y system of the centrifuge requires that you do not overload the rotor There are two options available for centrifuging samples whose weight including adaptor exceeds the maximum permissible load Redu...

Страница 26: ...mum number of cycles is given in the rotor table in the rotor specification section The maximum number of cycles for buckets is marked on the buckets themselves Service life examples nadm nmax Maximum...

Страница 27: ...Thermo Scientific 75003658 TX 200 5 1 5 Aerosol tight Applications Contents Basic Principles on page 5 2 Fill Level on page 5 2 Checking the Aerosol Tightness on page 5 2...

Страница 28: ...the rated level Checking the Aerosol Tightness CAUTION When centrifuging hazardous samples do not open aerosol tight rotors or buckets unless placed in a safety cabinet Always bear in mind the maximu...

Страница 29: ...ing your hands This releases the carbonic acid gas which is bound in the water resulting in excess pressure Do not apply pressure to the lid when doing so Leaks can be detected by escaping water or th...

Страница 30: ......

Страница 31: ...658 TX 200 6 1 6 Maintenance and Care Contents Cleaning intervals on page 6 2 Cleaning on page 6 2 Disinfection on page 6 3 Decontamination on page 6 4 Autoclaving on page 6 5 Service of Thermo Fisher...

Страница 32: ...arts Use only disinfectants with a pH of 6 8 Dry aluminium parts off with a soft cloth After cleaning treat the entire surface of aluminium parts with corrosion protection oil 70009824 Also treat the...

Страница 33: ...material has spilled during centrifugation Use a sprayer whenever possible so that all surfaces are covered evenly The rotor chamber and the rotor should be treated preferably with a neutral disinfec...

Страница 34: ...at a maximum temperature of 50 C 11 After cleaning treat the entire surface of aluminium parts with corrosion protection oil 70009824 Also treat the cavities with oil 12 Treat the bolt of the swing o...

Страница 35: ...tor with bolt grease 75003786 Autoclaving 1 Before autoclaving clean the rotor and accessories as described above 2 Place the rotor on a flat surface Rotors and adapter can be autoclaved at 121 C The...

Страница 36: ......

Страница 37: ...0 6 4 16 5 87 223 1200 6 4 16 5 103 266 1300 6 4 16 5 121 312 1400 6 4 16 5 140 362 1500 6 4 16 5 161 415 1600 6 4 16 5 183 472 1700 6 4 16 5 207 533 1800 6 4 16 5 232 598 1900 6 4 16 5 258 666 2000 6...

Страница 38: ...0 6 4 16 5 1088 2806 4000 6 4 16 5 1145 2952 4100 6 4 16 5 1203 3101 4200 6 4 16 5 1262 3254 4300 6 4 16 5 1323 3411 4400 6 4 16 5 1385 3571 4500 6 4 16 5 1449 3736 4600 6 4 16 5 1514 3903 4700 6 4 16...

Страница 39: ...S U S S S S M S S S S S S S S S M U U S S Formic Acid 100 S M U U U S M U U S S U S U S U Ammonium Acetate S S U S S S S S S S S S S U S S S S S S S S S S Ammonium Carbonate M S U S S S S S S S S S S...

Страница 40: ...U S U S U S S S M S S S S S S Dimethylsulfoxide S S U U S S S S U S S U S U U S S U U S S S S U U Dioxane M S U U S S M M S U U S U M U U M M M U S S S S U U Ferric Chloride U U S M S M S S S S M U S...

Страница 41: ...oxide conc U U M U M M S S U M U U U S M M U U U U Potassium Permanganate S S S S S S S S S U S S S M S M S U S S M S U S Calcium Chloride M U S S S S S S S S S S S S M S S S S S S S M S S S Calcium H...

Страница 42: ...S S S M S S S S S S M S S S S S S S S S S Nickel Salts U S S S S S S S S S S S S S S S S S S M S S S Oils Petroleum S S S S U S S S S U U M S M U U S S S U S S S S Oils Other S S S M S S S S U S S S...

Страница 43: ...Toluene S S U U S S M U S U U S U U U S U M U U U S U S U U M Trichloroacetic Acid U U U S S U M S U S U U S M M S S U U S U U U M U Trichloroethane S U M U U S U U U U U U U U U S U S S Trichloroethy...

Страница 44: ...or materials under the stress of centrifugation when in doubt we recommend pretesting sample lots Key S Satisfactory M Moderate attack may be satisfactory for use in centrifuge depending on length of...

Страница 45: ...ity Chart B 1 Cleaning 6 2 Cleaning agents 6 3 Cycle Counter 4 4 D Decontamination 6 4 Disinfection 6 3 F Fill Level 5 2 I Improper Loading 4 3 M Maintenance 6 1 Maximum Loading 4 3 P Placing O ring 5...

Страница 46: ......

Страница 47: ...erms and pricing are subject to change Not all products are available in all countries Please consult your local sales representative for details Shown pictures within the manual are examples and may...

Отзывы: