C-Link Protocol Commands
Communications Configuration
Thermo Fisher Scientific
Model 49
i
Primary Standard Instruction Manual
B-39
Receive:
set host name analyzer01 ok
instr name
This command reports the instrument name.
Send:
instr name
Receive:
instr name
o3 Analyzer
03 Analyzer
instrument id
This command reports the instrument id.
Send:
instrument id
Receive:
instrument id 49
set instrument id
value
This command sets the instrument id to
value,
where
value
is a decimal
number between 0 and 127 inclusive.
Note
Sending this command via RS-232 or RS-485 will require the host to
use the new id for subsequent commands.
Send:
set instrument id 50
Receive:
set instrument id 50 ok
mode
This command reports what operating mode the instrument is in: local,
service, or remote. The example that follows shows that the instrument is in
the remote mode.
Send:
mode
Receive:
mode remote
set mode local
set mode remote
These commands set the instrument to local or remote mode. The example
that follows sets the instrument to the local mode.
Send:
set mode local
Receive:
set mode local ok
program no
This command reports the analyzer’s model information and program
version number, which will be dependant on the current version.
Содержание 49i
Страница 4: ......
Страница 12: ...Figures xii Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 146: ...Troubleshooting Service Locations 5 22 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 203: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 5 Figure 8 2 Bench Mounting ...
Страница 204: ...8 6 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific Figure 8 3 EIA Rack Mounting ...
Страница 205: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 7 Figure 8 4 Retrofit Rack Mounting ...
Страница 206: ...Optional Equipment Mounting Options 8 8 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...