![Thermo 42i Скачать руководство пользователя страница 290](http://html.mh-extra.com/html/thermo/42i/42i_instruction-manual_1104217290.webp)
C-Link Protocol Commands
I/O Configuration
B-42
Model 42
i
Low Source Instruction Manual
Thermo Electron Corporation
dtoa
channel
This reports the outputs of the 6 or 12 digital to analog converters,
according to Table B-12. The example below shows that the DAC 1 is
97.7% full-scale.
Send:
dtoa 1
Receive:
dtoa 1 97.7%
Note
All channel ranges are user definable. If any customization has been
made to the analog output configuration, the default selections may not
apply.
list var aout
list var dout
list var din
These commands report the list of index numbers, and the variables
(associated with that index number) available for selection in the current
mode (determined by single/dual/auto, gas mode no/no
x
or no/no
x
) for
analog output, digital output and digital inputs. The index number is used
to insert the variable in a field location in a list using “set sp
field index
”. The
example below reports the list of analog output, index numbers, and
variables.
Table B-12.
Default Output Assignment
D to A
Function
Single Range
Dual Range
Auto Range
1
Voltage Output
NO
High NO
High/Low NO
2
Voltage Output
NO
2
High NO
2
High/Low NO
2
3
Voltage Output
NO
x
High NO
x
High/Low NO
x
4
Voltage Output
Not Assigned
Low NO
Range Status
5
Voltage Output
Not Assigned
Low NO
2
Not Assigned
6
Voltage Output
Not Assigned
Low NO
x
Not Assigned
7
Current Output
NO
High NO
High/Low NO
8
Current Output
NO
2
High NO
2
High/Low NO
2
9
Current Output
NO
x
High NO
x
High/Low NO
x
10
Current Output
Not Assigned
Low NO
Range Status
11
Current Output
Not Assigned
Low NO
2
Not Assigned
12
Current Output
Not Assigned
Low NO
x
Not Assigned
Содержание 42i
Страница 4: ...WEEE Compliance WEEE Compliance Thermo Electron Corporation ...
Страница 14: ...Tables xiv Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 28: ...Installation Startup 2 6 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 128: ...Operation Password 3 100 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 146: ...Calibration Zero and Span Check 4 18 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 154: ...Preventive Maintenance Pump Rebuilding 5 8 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 224: ...Servicing Service Locations 7 48 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 238: ...System Description I O Components 8 14 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 246: ...Optional Equipment Mounting Options 9 8 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 314: ......