Servicing
Ozonator Assembly Replacement
Thermo Electron Corporation
Model 42
i
Low Source Instruction Manual
7-25
3. Loosen the four captive screws securing the ozonator bracket to the
floor plate.
Figure 7-14.
Replacing the Ozonator Assembly
4. Remove the two screws securing the ozonator to the ozonator bracket.
5. Unplug the ozonator from the ozonator transformer by lifting the
ozonator straight up.
6. To install the ozonator, follow the previous steps in reverse.
7. Re-install the measurement bench. Refer to “Removing the
Measurement Bench and Lowering the Partition Panel” in this chapter.
Ozonator Assembly - Top View
Ozonator Assembly - Side View
Captive Screws (4)
Ozonator Bracket Screws (2)
Ozonator
Ozonator Transformer
Screws (4)
Ozonator Transformer
Содержание 42i
Страница 4: ...WEEE Compliance WEEE Compliance Thermo Electron Corporation ...
Страница 14: ...Tables xiv Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 28: ...Installation Startup 2 6 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 128: ...Operation Password 3 100 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 146: ...Calibration Zero and Span Check 4 18 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 154: ...Preventive Maintenance Pump Rebuilding 5 8 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 224: ...Servicing Service Locations 7 48 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 238: ...System Description I O Components 8 14 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 246: ...Optional Equipment Mounting Options 9 8 Model 42i Low Source Instruction Manual Thermo Electron Corporation ...
Страница 314: ......