5
•
Installation
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Page 71
5.6.3
Connecting Fittings, Capillaries, and Tubing
This section provides information about how to connect and handle
capillaries, fittings, and tubing.
5.6.3.1
General Guidelines
When connecting capillaries and tubing, follow these general
recommendations:
•
Use only the capillaries and tubing (for example, solvent lines or
waste tubing) that are shipped with the device or additional or spare
capillaries and tubing as recommended by Thermo Fisher Scientific.
•
The connectors must be free from contaminants. Even minor
particles may cause damage to the system or lead to invalid test
results.
•
Do not install capillaries or tubing that is stressed, nicked, kinked, or
otherwise damaged.
•
Install capillaries and fittings only at the positions for which they are
intended.
5.6.3.2
Overall Flow Path Connection
A delay capillary is used to connect HPLC detector outlet and fraction
collection valve central port (Inlet).
A detector waste line is shipped with HPLC detector. However, if a
fraction collector is installed in the HPLC system, the waste line will be
used to connect with the Y-connector outlet of the fraction collector and
routed to waste reservoir.
Содержание VF-F20-A
Страница 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Страница 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Страница 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Страница 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Страница 231: ...10 Accessories Consumables and Replacement Parts Integral Fraction Collector FT VF F20 A Operating Manual Page 231 ...
Страница 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...
Страница 240: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...