6
•
Operation
Integral Fraction Collector FT (VF-F20-A) Operating Manual
Page 139
•
Make sure that the temperature of the column does not exceed
40 °C.
•
When resuming operation, let the flow equilibrate and verify that
the operating parameters for the other system modules are set as
required before proceeding.
6.10.2
Long-Term Shutdown
Shutting Down the device
TIP
Shutting down the device affects the operation of the system. When
shutting down the device, also observe the shutting down instructions
for the other Vanquish system modules and take appropriate action
(refer to the
Operating Manuals
for the modules).
To interrupt operation for a longer period, follow the instructions below.
1. Check that the fraction collection valve is set to Collect position.
2. Flush the system with an appropriate, pure solvent (minimum HPLC-
grade).
Observe the following:
Situation after Shutdown
If no additive is used
If an addititive is used
Device remains in the
laboratory after shutdown
Flush the system, for
example with methanol.
100% acetonitrile should
not be used.
Flush the system with
several volumes (for
example, 1.0 mL/min for 10
minutes with the standard
system) of methanol and
water (50:50) to prevent
salt buildup in the fluidics. If
the solvents in the device
are not miscible with water,
use an appropriate
intermediate solvent.
Содержание VF-F20-A
Страница 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Страница 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Страница 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Страница 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Страница 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Страница 231: ...10 Accessories Consumables and Replacement Parts Integral Fraction Collector FT VF F20 A Operating Manual Page 231 ...
Страница 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...
Страница 240: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...