6
•
Operation
Page 98
Vanquish Pumps (VH-P10-A-02) Operating Manual
6.8
Purging the Pump
Purging the pump means rinsing the pump for a short time at a higher
flow rate, using the selected solvents. From the purge unit, the solvent is
automatically dispensed to waste.
When
•
The pump is operated for the first time
To flush out the isopropanol that is present in some components of
the pump when the pump is shipped from the manufacturing site
•
One or more solvent lines have been empty
•
To remove air bubbles that might be trapped in the system (pump
head, solvent lines) because air bubbles in the system may lead to
the following problems:
♦
Pressure pulsation
♦
High noise level or pulsation during pump operation
♦
Non-reproducible analysis
•
When changing to a different type of solvent, for example, when
changing to a different analytical method
Items required
Solvents suitable for purging
To flush out the isopropanol when operating the pump for the first time,
use solvents that are miscible with isopropanol. If they are not, use an
appropriate intermediate solvent.
Settings
The following settings are considered for the purge cycle:
Setting
Description
Purge flow
The standard setting is 5 mL/min.
You can adapt the setting in Chromeleon if required. The setting
is considered also when purging the pump from the keypad.
Purge time
The standard purge time is 5 minutes.
You can adapt the setting in Chromeleon if required. The setting
is considered also when purging the pump from the keypad.
Содержание Vanquish VH-P10-A-02
Страница 1: ...Vanquish Pumps VH P10 A 02 Operating Manual 4820 4411 EN Revision 1 0 October 2017 thermoscientific ...
Страница 4: ...Contacting Us Page 4 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 10: ...Contents Page 10 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 30: ...2 Safety Page 30 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 46: ...4 Unpacking Page 46 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 80: ...5 Installation Page 80 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 232: ...11 Appendix Page 232 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 233: ...12 Index Vanquish Pumps VH P10 A 02 Operating Manual Page 233 12 Index ...
Страница 238: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...