6
•
Operation
Page 112
Vanquish Pumps (VH-P10-A-02) Operating Manual
6.10.2
Long-Term Shutdown
Shutting Down the Pump
To interrupt operation for a longer period, follow the instructions in this
section.
TIP
Shutting down the pump affects the operation of the system. When
shutting down the pump, also observe the shutting down instructions
for the other Vanquish system modules and take appropriate action
(refer to the
Operating Manuals
for the modules).
1. Remove the column.
2. Flush the system with an appropriate, pure solvent (minimum HPLC-
grade).
Observe the following:
♦
Pump remains in the laboratory after shutdown
♦
If no additive is used, flush the system, for example with
methanol. 100% acetonitrile should not be used.
♦
If an additive is used, flush the system with several volumes
(for example, 1.0 mL/min for 10 minutes with the standard
system) of methanol and water (50:50) to prevent salt
buildup in the fluidics. If the solvents in the pump are not
miscible with water, use an appropriate intermediate
solvent.
♦
Pump shall be transported or shipped after shutdown
♦
If no additive is used, flush the system with isopropanol.
♦
If an additive is used, flush the system first with several
volumes (for example, 1.0 mL/min for 10 minutes with the
standard system) of methanol and water (50:50) to prevent
salt buildup in the fluidics. If the solvents in the pump are
not miscible with water, use an appropriate intermediate
solvent. Afterward, flush the system with isopropanol.
3. Turn off the pump flow and wait until the system pressure is down
to zero.
Содержание Vanquish VH-P10-A-02
Страница 1: ...Vanquish Pumps VH P10 A 02 Operating Manual 4820 4411 EN Revision 1 0 October 2017 thermoscientific ...
Страница 4: ...Contacting Us Page 4 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 10: ...Contents Page 10 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 30: ...2 Safety Page 30 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 46: ...4 Unpacking Page 46 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 80: ...5 Installation Page 80 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 232: ...11 Appendix Page 232 Vanquish Pumps VH P10 A 02 Operating Manual ...
Страница 233: ...12 Index Vanquish Pumps VH P10 A 02 Operating Manual Page 233 12 Index ...
Страница 238: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com ...