6
•
Operation
Page 100
Vanquish Charged Aerosol Detectors Operating Manual
6.6
Important Operating Parameters
The parameters described in this section should be considered for
routine operation of the detector. You can usually access these
parameters from the Chromeleon user interface. For more information,
refer to
Chromeleon Help and documents
.
Parameters
Description
Signal channels
The signal channels are available for the detector in the
Instrument Configuration Manager (Chromeleon 7) or
Server Configuration program (Chromeleon 6.8).
CAD_1
The
CAD_1
signal channel is available for data acquisition.
As a standard, the detector measures the current in
pA
.
Background current
Records and monitors the background current of the
detector during data acquisition without autozero or filter
constant applied. As a standard, the detector measures
the current in
pA
.
For further information, see
section 6.7.4 Signal Channels,
page 106
Gas regulator
pressure (signal
channel)
Records and monitors the nebulizer pressure during data
acquisition.
The default unit setting for the gas pressure corresponds
to the
Pressure Unit
that is defined on the detector
configuration pages in the Chromeleon software.
For further information, see
section 6.7.4 Signal Channels,
page 106
Evaporation tube
temperature
Records and monitors the actual temperature of the
evaporation tube during data acquisition. The
temperature is recorded in
°C
as a default.
For further information, see
section 6.7.4 Signal Channels,
page 106
Gas state
Turns the detector gas flow
On
or
Off
.
Observe the notes in
section 6.5.3 Turning On the
Autozero
Performs an automatic null balancing. The current
detector signal is interpreted as 0 at the current range in
use.
Содержание Vanquish VF-D20
Страница 4: ...Contacting Us Page 4 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 28: ...2 Safety Page 28 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 38: ...3 Detector Overview Page 38 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 44: ...4 Unpacking and Transport Page 44 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 150: ...7 Maintenance and Service Page 150 Vanquish Charged Aerosol Detectors Operating Manual...
Страница 177: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 177 12 Index...
Страница 181: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 181...
Страница 182: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...