5
•
Maintenance and Service
Solvent Monitor
Installation Guide
Page 33
5.5
Replacing the Retaining Guides
When
Retaining guide is damaged.
Parts required
Replacement retaining guides
Preparations
1. Turn off the pump flow and wait until the system pressure is down
to zero.
Follow these steps
1. Unscrew the cap of the solvent reservoir/waste container and
remove the solvent monitor line together with the cap from the
solvent reservoir/waste container.
2. Remove the end cap from the solvent monitor line.
3. Pull the solvent monitor line out of the solvent reservoir/waste
container cap.
4. Remove the retaining guide.
5. Slip a new retaining guide onto the solvent monitor line.
6. Feed the dipstick of solvent monitor line through an open hole in a
solvent reservoir/waste container cap.
7. Attach an end cap to the dipstick of the solvent monitor line.
8. Insert the dipstick of the solvent monitor line with the end cap into
the solvent reservoir/waste container and tighten the cap onto the
solvent reservoir/waste container.
9. Adjust the solvent monitor line so that the end cap rests on the
bottom of the solvent reservoir/waste container and that the
dipstick is straight.
10. Press the retaining guide into the cap.
11. To remove any air bubbles from the solvent lines, purge the pump.