Universal Chromatography Interfaces
UCI-50 and UCI-100
Operating Instructions
Page
27
3.3.5 Setting Up the UCI in Chromeleon
All software details in this manual refer to
Chromeleon 6.80
. If you want to operate the
UCI with Chromeleon 7, note the information on page 10.
3.3.5.1 Installing the UCI
After the USB software for the NCS has been installed (
→
page 13) or the LAN connection
has been established (
→
page 15), install and configure the UCI in Chromeleon:
1.
Start the Chromeleon
Server Monitor
and the Chromeleon server if they are not yet
running.
a)
Open the
Chromeleon Server Monitor
program by double-clicking the Chromeleon
Server Monitor icon
on the Windows taskbar. (If the Server Monitor icon is not
on the taskbar, click
Start
on the taskbar, point to
Programs
(or
All Programs
,
depending on the operating system), point to
Chromeleon
, and then click
Server
Monitor
.)
b)
Click
Start
to start the server.
c)
Click
Close
to close the Server Monitor window. The Server Monitor icon
appears on the taskbar. (Clicking the
Quit Monitor
button quits (exits) the Server
Monitor program, but does not stop the server. To stop the server, click
Stop
.)
2.
Start the Chromeleon
Server Configuration
program by clicking
Start
on the taskbar.
Point to
Programs
(or
All Programs
, depending on the operating system), point to
Chromeleon
, and then click
Server Configuration
.
3.
Click the plus sign next to the server name to display the items underneath.
4.
Right-click
Sharable Devices
in the timebase and click
Add Sharable Device
on the
menu.
5.
In the list box, select
Dionex UCI-1000 Interface
(or
UCI-50 Interface
) and click
OK
to confirm the selection.
6.
The configuration pages for the UCI are opened. On each page, verify that the settings
are correct and select additional settings if needed. For a description of the pages, see
section 3.3.5.2 (
→
page 28).
7.
On the
File
menu, click
Save Installation
and then close the Server Configuration
program.
Содержание UCI-100
Страница 2: ...Universal Chromatography Interfaces UCI 50 and UCI 100 Operating Instructions...
Страница 4: ...Universal Chromatography Interfaces UCI 50 and UCI 100 Page II Operating Instructions...
Страница 10: ...Universal Chromatography Interfaces UCI 50 and UCI 100 Page 4 Operating Instructions...
Страница 48: ...Universal Chromatography Interfaces UCI 50 and UCI 100 Page 42 Operating Instructions...
Страница 52: ...Universal Chromatography Interfaces UCI 50 and UCI 100 Page 46 Operating Instructions...
Страница 54: ...Universal Chromatography Interfaces UCI 50 and UCI 100 Page 48 Operating Instructions...
Страница 60: ...Universal Chromatography Interfaces UCI 50 and UCI 100 Page 54 Operating Instructions...