5
Changing Ionization Modes
Removing the Ion Volume
Thermo Scientific
TSQ Quantum XLS Series User Guide
77
Figure 46.
I/R tool at first stop and inlet valve lever up
The inlet valve is now open.
8. Remove the ion volume:
a. Slide the I/R tool into the vacuum manifold until the tip of the I/R tool is fully
inserted into the ion volume holder. See
. The head of the arrow on the I/R
tool, when viewed through the window of the ion source manifold, should not be
visible.
Inlet valve lever
First stop of
guide bar
I/R tool