Maintenance
Removing and Installing a GC Capillary Column
110
TSQ Quantum XLS Series User Guide
Thermo Scientific
f.
Allow the column
to purge for at least 10 minutes.
3. Perform a manual column leak check. For instructions, refer to the Ensuring Tightness
chapter of the
TRACE GC Ultra Maintenance and Troubleshooting Manual
.
4. Perform a column evaluation. For instructions, refer to the Columns chapter of the
TRACE GC Ultra Operating Manual
. For expected K factors for various column
diameters, column lengths, and carrier gases refer to the
K-Factor Quick Reference
card.
5. Perform an automatic column leak check:
a. Run an automatic leak check. Refer to the Columns chapter of the
TRACE GC Ultra
Operating Manual
for instructions.
b. If the report indicates a leak, look for leaks and use the leak detector to fix leaks at all
the fittings in the GC.
c.
Repeat column evaluation and leak check procedures until no leaks can be found.
6. Condition the capillary column according to the manufacturer’s instructions. You must
condition new columns before inserting them in the mass spectrometer.
7. Connect the column to the transfer line:
a. Shut down and vent the mass spectrometer. See
b. Lower the GC oven temperature to 30 °C and allow it to cool before continuing.
Note
The K factor is a measure of the resistance of the column. A K factor that is too
low might indicate a leak in the system, while a K factor that is too high might
indicate a blockage.
CAUTION
Do not raise the oven temperature until you are sure the system is
leak-free. At temperatures above 100 °C, oxygen exposure will destroy the
column.
CAUTION
Do not insert the column into the transfer line during the conditioning.
The material released from the column during the conditioning (column bleed) will
contaminate the ion source, which you must then clean.
CAUTION
Never exceed the column manufacturer’s maximum operating
temperature.
WARNING
Burn Hazard
. The GC oven and transfer line might be hot. Allow
them to cool to room temperature before touching them. Do not touch the
injector when it is hot.
Содержание TSQ Quantum XLS
Страница 1: ...TSQ Series TSQ Quantum XLS Series User Guide 70111 97225 Revision A December 2011 ...
Страница 3: ......
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 32: ......
Страница 64: ......
Страница 76: ......
Страница 102: ...86 TSQ Quantum XLS Series User Guide Thermo Scientific Figure 50 Ion source assembly assembled and exploded views ...
Страница 160: ......
Страница 166: ......
Страница 176: ......