Maintenance
Replacing the Filament of the TSQ Quantum GC Ion Source
106
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
2. Shut down and vent the TSQ Quantum GC. See
3. Remove the ion source. See
“Removing the TSQ Quantum GC Ion Source”
4. Wearing clean, lint- and powder- free gloves, remove the ion source from the magnet
yoke.
Figure 57.
Disassembling the ion source to replace the filament
5. Unscrew the threaded standoff and remove the retainer clip that secures the heater ring
and lens assembly to the ion source block. See
6. Pull the heater ring and lens assembly out of the ion source block.
7. Remove the retainer clip that secures the filament and centering ring to the ion source
block.
CAUTION
Shock Hazard
. Unplug the TSQ Quantum GC before proceeding.
Filament retainer clip
Filament
Centering ring
Heater ring and lens
assembly
Heater ring and lens
assembly retainer clip
Threaded standoff
Ion source block and
EI/CI Source PCB
Filament retainer clip
Filament
Centering ring