16
Optimizing the Compound Signal
Optimizing Parameters in the Optimization Pane
68
TSQ Altis, Quantis, and Fortis Hardware Manual
Thermo Scientific
Optimizing Parameters in the Optimization Pane
1. Make sure that the syringe has enough of the sample solution.
2. In the Tune window, click
Syringe Off
to turn on the syringe pump.
The button name changes to Syringe On. If your LC/MS system includes an LC device,
prepare an LC method before you turn on the liquid flow to the API source.
3. In the Optimization pane (
), do the following:
a. In the Sample Injection Type list, select
Syringe
.
b. In the Mass List Type list, make a selection.
If you select the Formula mass list type, you can select one or both ion polarity modes
and specify the adducts to include. For descriptions of the Optimization parameters,
refer to the Tune Help.
c.
In the Source Optimization list, select
On
and then select the
Spray Voltage
,
Sheath
Gas
,
Aux Gas
, and
Sweep Gas
check boxes that appear.
d. For a SIM scan, select the
Source Fragmentation
,
RF Lens
, and
Precursor Ion Mass
check boxes. For an SRM scan, also select
On
in the Product Ion Optimization list.
e.
Click the
Add/Remove Table Column
icon,
, next to the parameters that you
want to customize per compound. Otherwise, the MS applies the parameter setting
to all compounds.
When selected, the icon changes color, from
to
, and the parameter appears in
the Optimization Table (not shown here).
IMPORTANT
Thermo Fisher Scientific recommends that you optimize system parameters
if the spray becomes unstable and also after restarting the sample flow into the system.
Note
The default values are sufficient for most experiments.
Содержание TSQ Altis
Страница 1: ...TSQ Series II TSQ Altis TSQ Quantis and TSQ Fortis Hardware Manual 80111 97047 Revision A June 2018 ...
Страница 6: ......
Страница 12: ...Contents xii TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 22: ...1 Introduction MS Mass To Charge Ratio Ranges 4 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 32: ...3 Vacuum System Vacuum Gauges 14 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 172: ...G 154 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...