Using Basic Tune Functions
Setting the Instrument Power Mode
116
TSQ Altis, Quantis, and Fortis Hardware Manual
Thermo Scientific
Setting the Instrument Power Mode
Use the three power mode icons in the Tune window (
) to set the MS’s power mode
(on, standby, or off ).
When you remove the API source housing or the spray insert, the MS automatically switches
to off mode.
In standby mode, the System LED on the front panel turns yellow and the MS turns off the
,
, 8 kV power to the API source, main RF voltage,
and ion optic RF voltages. The auxiliary, sheath, and sweep gas flows remain on and return to
their standby default settings (2 arbitrary). See
Chapter 8, “Using the MS Power Options.”
Divert/inject valve
To set the divert/inject valve module and position
Click the down arrow to select the valve, and then
click
Valve
to set the position.
Syringe pump
To turn the syringe pump on or off
Click
Syringe On (Off )
to switch between on and
off.
To set the syringe pump parameters
1. Click the down arrow, and then type the values.
The Tune application automatically saves the values.
2. (Optional) Press and hold
Prime
to prime the syringe
at 100 μL/min.
3. Click the down arrow again or click elsewhere to
close the box.
Table 17.
Procedures for using the instrument control buttons (Sheet 2 of 2)
Button
Function
Procedure
Содержание TSQ Altis
Страница 1: ...TSQ Series II TSQ Altis TSQ Quantis and TSQ Fortis Hardware Manual 80111 97047 Revision A June 2018 ...
Страница 6: ......
Страница 12: ...Contents xii TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 22: ...1 Introduction MS Mass To Charge Ratio Ranges 4 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 32: ...3 Vacuum System Vacuum Gauges 14 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 172: ...G 154 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...