19
Acquiring Sample Data
Using Tune to Acquire Sample Data
82
TSQ Altis, Quantis, and Fortis Hardware Manual
Thermo Scientific
Using Tune to Acquire Sample Data
1. Open the Data Acquisition pane (
), and then do the following:
a. (Optional) To change the destination folder for the raw data file, click the
Browse
icon.
The default folder location is in
drive
:\Thermo\Data.
b. In the File Name box, type
reserpine
(or the name of the analyte).
If the base file name already exists in the save location, the Tune application adds a
time-stamp suffix that consists of the year (
YYYY
), month (
MM
), day (
DD
), and
time (
HHMMSS
).
c.
In the Sample Name box, type the name of the analyte (or other suitable label).
d. In the Comment box, type a comment about the experiment.
For example, describe the ionization mode, scan type, scan rate, sample amount, or
method of sample introduction. The data system includes the comment in the header
information for the raw data file.
You can also add this information to reports created with the Xcalibur XReport
reporting application. To open this application, choose
Start > All Programs >
Thermo Xcalibur > XReport
.
e.
Under Timed Acquisition, select the
Continuous Acquisition
option from the drop
down list (acquires data until you stop the acquisition)(1) in (
).
Содержание TSQ Altis
Страница 1: ...TSQ Series II TSQ Altis TSQ Quantis and TSQ Fortis Hardware Manual 80111 97047 Revision A June 2018 ...
Страница 6: ......
Страница 12: ...Contents xii TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 22: ...1 Introduction MS Mass To Charge Ratio Ranges 4 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 32: ...3 Vacuum System Vacuum Gauges 14 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...
Страница 172: ...G 154 TSQ Altis Quantis and Fortis Hardware Manual Thermo Scientific ...