1
Performing Routine Maintenance
Maintaining the Calibration Gas Module
Thermo Scientific
TSQ 9000 Hardware Manual
71
4. Inject the FC 43. If you see liquid pooled on top of the white frit, remove the excess
liquid according to local environmental regulations.
5. Reattach the cover to the calibration gas controller.
6. Close the front door of the instrument.
CAUTION INSTRUMENT DAMAGE:
Adding more than 300 μL of calibration compound
can damage the calibration gas controller. Be sure liquid does not get into the controller
when you reattach the reservoir.
ATTENTION DOMMAGES À L’INSTRUMENT
: ajouter plus de 300 μl de composé
d’étalonnage à la fois peut endommager le module de gaz d’étalonnage. Assurez-vous
qu’aucun liquide n’entre dans le module lorsque vous rattachez le réservoir.