![Thermo Scientific TSQ 9000 Скачать руководство пользователя страница 430](http://html.mh-extra.com/html/thermo-scientific/tsq-9000/tsq-9000_hardware-manual_1104856430.webp)
3
Advanced Troubleshooting
Replacing Fans and Filters
402
TSQ 9000 Hardware Manual
Thermo Scientific
Turbo Cooling Fan and Turbo Cooling Fan Filter
To replace the turbo cooling fan and turbo cooling fan filter
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
. You do not need to remove the ion source cartridge.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. Move the GC away from the TSQ 9000 instrument.
4. Use a T20 Torxhead screwdriver to remove the five screws on the right side panel.
5. Lift the right side panel off the instrument.
6. Use a T20 Torxhead screwdriver to remove the four screws securing the left hand front
panel to the chassis. One of the screws is located on the interior front panel of the
instrument.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for ordering information.
Figure 427.
Removing the Right Side Panel Screws
Right Side
Panel Screws