Advanced Troubleshooting
Replacing the Calibration Gas Components
Thermo Scientific
TSQ 9000 Hardware Manual
377
5. Attach a new calibrant reservoir to the calibration gas module.
6. Fill the calibrant reservoir of the new module by following the instructions in
7. Reattach the calibrant reservoir cover.
8. Close the front door of the instrument.
9. If you are using hydrogen as a carrier gas, attach the hydrogen safety screws and restart
your GC by following the instructions in
Figure 401.
Replacing the Calibrant Reservoir and Cover
Calibrant
Reservoir
Cover
Calibrant
Reservoir