Advanced Troubleshooting
Replacing the Power Supplies
362
TSQ 9000 Hardware Manual
Thermo Scientific
Dynode and Multiplier Power Supply
To replace the dynode and multiplier power supply
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. Remove all power connections from the TSQ 9000 instrument.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
You do not have to remove the ion source cartridge or column or vent the
instrument. If you are using hydrogen as a carrier gas, power off the GC.