14. From the left side of the instrument, use a T20 Torxhead screwdriver to remove the three
screws around the 20-pin feedthrough.
15. Remove and replace the 20-pin feedthrough. If necessary, replace the o-ring and make
sure it is properly aligned and free of dust.
Figure 351.
Removing the 20-Pin Feedthrough
Figure 352.
Replacing the 20-Pin Feedthrough
20-Pin Feedthrough
Screws
O-ring