Advanced Troubleshooting
Replacing the Covers of the TSQ 9000 Instrument
296
TSQ 9000 Hardware Manual
Thermo Scientific
Top Manifold Cover
To replace the top manifold cover
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
. You do not have to remove the ion source cartridge or column. However,
you must vent the instrument. If you are using hydrogen as a carrier gas, power off the
GC.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. Remove and replace the large or small glass manifold covers as needed on the top of the
instrument.
4. Replace the top cover panel and tighten the screw holding it in place.
5. Restart your system by following the instructions in
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Figure 317.
Replacing the Small and Large Glass Manifold Covers
Small Glass Manifold Cover
Large Glass Manifold Cover