Advanced Troubleshooting
Replacing the Covers of the TSQ 9000 Instrument
Thermo Scientific
TSQ 9000 Hardware Manual
293
Left Hand Front Panel and Left Hand Sub Panel
To replace the left hand front panel and the left hand sub panel of the TSQ 9000
instrument
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
. You do not have to remove the ion source cartridge or column or vent the
system. If you are using hydrogen as a carrier gas, power off the GC.
2. Remove the four screws on the left hand front panel. One is located on the bottom left of
the interior front panel of the instrument.
Figure 314.
Locating the Left Hand Front Panel Screws
3. To replace the left hand sub panel, disconnect the inputs as shown in
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Screws
Screw on Front