Advanced Troubleshooting
Replacing Components of the Electron Multiplier
Thermo Scientific
TSQ 9000 Hardware Manual
283
Anode Feedthrough
To replace the anode feedthrough, o-ring, four screws, and a standoff:
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. Disconnect the electrometer board cable and large feedthrough cables labeled
Dynode
and
Electron Multiplier
.
4. Lift the electrometer shield off the instrument and set it aside.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
You do not have to remove the ion source cartridge or the column. However, you
must vent the instrument.
Figure 307.
Disconnecting the Electron Multiplier Cables
Electron
Multiplier Plate
Dynode
Electrometer
Board Cable
Electron
Multiplier