Advanced Troubleshooting
Replacing Components of the Ion Source
192
TSQ 9000 Hardware Manual
Thermo Scientific
Source to Ion Guide Spacer
To replace the source to ion guide spacer:
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
.
2. Remove the analyzer tray from the instrument by following the instructions in
“Removing the Analyzer Tray from the TSQ 9000 Instrument”
3. Disconnect the Q1 entrance lens, Q1 exit lens, Q3 entrance lens, and Q3 exit lens from
the source interface board using a flat-tipped screwdriver.
Figure 175.
Locating Wires on the Source Interface Board
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
WARNING - INSTRUMENT DAMAGE
Thermo Fisher Scientific strongly recommends
that only trained Thermo Fisher Scientific Field Service Engineers remove and service this
component.