3
Advanced Troubleshooting
Replacing Components of the Ion Source
190
TSQ 9000 Hardware Manual
Thermo Scientific
Lens Plate and Springs
To replace the lens plate and springs
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
.
2. Remove the analyzer tray from the instrument by following the instructions in
“Removing the Analyzer Tray from the TSQ 9000 Instrument”
3. Use a T10 Torxhead screwdriver to disconnect the lens plate and springs from the lens
heater block.
4. Disconnect the lens 1, lens 2, and lens 3 wires from the source interface board using a
flat-tipped screwdriver.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering
these components.
Figure 173.
Disconnecting the Lens Heater Block Grounding Strap
Lens
Heater
Block
Lens Plate and
Springs
Screw