Troubleshooting
Investigating RF/DC Issues
Thermo Scientific
TSQ 9000 Hardware Manual
117
Investigating RF/DC Issues
When you run RF/DC System Check diagnostics on the TSQ 9000 system, the Xcalibur
software checks the RF/DC system of the instrument. Your instrument or software may
indicate the following types of issues.
RF dip is incorrect.
RF/DC System Check diagnostic fails.
WARNING
-
ELECTRICAL SHOCK HAZARD:
When troubleshooting any
issue that requires removing a cover on the TSQ 9000 mass spectrometer,
you should power-off and vent the instrument to avoid harming yourself.
AVERTISSEMENT RISQUE D’ÉLECTROCUTION
: lorsque vous cherchez à
résoudre un problème nécessitant d’enlever le couvercle du spectromètre de
masse TSQ 9000 mettez hors tension et ventilez l’instrument pour éviter de
vous blesser.
Table 27.
Possible Remedy if the RF Dip is Incorrect
Make sure the RF board is properly connected and use
Xcalibur
Power Status to ensure the readback r48 V DC is present. If +48 V
DC is not present, contact Technical Support.
Table 28.
Possible Remedy if RF/DC System Check Diagnostic Fails
Use TSQ 9000 Tune to tune the RF system. If the RF system will not tune, contact Technical Support.
Make sure the rod driver board is properly installed and undamaged. Replace it, if necessary. See