5
Optimizing Your Method
Modifying an Automatic Tune
110
TSQ 8000 Evo User Guide
Thermo Scientific
Figure 107.
Configuring the Q3 Resolution
a.
Mass
—Click in this box to select the ion to be used for tuning.
b.
Peak Width
—Enter the target peak width.
c.
Measure at %
—Click in this box to select the location on the peak at which you
want to measure the target peak width.
•
10
—Measures the width at 10% of the peak height.
•
50
—Measures the width at 50% of the peak height.
d.
Lens Tune Relation
—Click in this box to set the occurrence of the resolution tune
parameters before or after a lens tune.
•
Before and After
—Uses the same resolution parameters before and after a lens
tune.
•
Before
—Uses the resolution parameters before a lens tune.
•
After
—Uses the resolution parameters after a lens tune.
5.
Tune High Scan Rate
—Select this check box to tune the resolution at a scan rate in
addition to the 100 and 1000 amu/s default scan rates.