Using the Solution Kit (Type H)
30
TruNarc User Guide
Thermo Scientific
Solution Kit
Preparing for a Solution Kit (Type H) Scan
Obtain an ethanol vial and a test stick package from the Solution Kit. Check
the expiration date on the package and on the label that is on the vial, and do
not use them if the date has passed.
When you are ready to do the scan, open a vial and a test stick package. Do
not remove the test stick from the package until you are ready to use it, to
protect it from contaminants. The test stick must be uncontaminated and
dry for the scan.
Performing a Solution Kit (Type H) Scan
This section explains how to scan narcotics using the Solution Kit (Type H).
• For the list of narcotics that can be identified using the Solution Kit
(Type H), reference the most recent library list.
• For information on scanning black tar heroin, see
Step 1
Remove the test stick from its package and hold it by the end opposite from
the scoop.
Fill up the scoop on the end of the test stick with the sample material. You
do not need to pack the sample material into the scoop.