1
TriPlus 300 Headspace Site Preparation
Gas Equipment Requirements
Thermo Scientific
TriPlus 300 Headspace Preinstallation Requirements Guide
7
2. Ensure that the relative humidity in your laboratory is between 40 and 75%, with no
condensation. However, the instrument perform best at 20-25 °C room temperature and
relative humidity of 40 to 50%. A temperature and humidity monitor in your laboratory
helps ensure that the climate is within these specifications.
3. Ensure that the air in your site is free of excess
particulate matter
.
For reference, the air should contain fewer than 100,000 particles (larger than 5
μ
m) per
cubic meter. If the concentration is larger than this amount, dust can accumulate on
electronic components. This accumulation reduces their ability to cool off properly and
could cause them to overheat. If your environment is particularly dusty, we recommend
that you purchase the optional dust filter for your system.
4. Ensure that your site is free of electrostatic discharge (ESD), which might damage the
electronic components of your system. Ensure your static has been discharged before
touching internal components of the instrument. ESD can damage sensitive components,
resulting in premature failures.
Take the following precaution to prevent electrostatic discharge:
• Use a static-dissipating floor covering (such as tile or conductive linoleum) in the
room housing your instrument.
• Use laboratory chairs covered with natural fibers or other static-dissipating material.
• Wear laboratory coats and clothing made from natural fibers or other
static-dissipating material.
• Do not place polystyrene (foam) cups or packing materials on the instrument.
Gas Equipment Requirements
Use the following guidelines to make sure you have the gas supplies for your system ready far
in advance of installation. You will need a supply of ultra-high purity GC gases.
The TriPlus 300 Headspace sampling system requires auxiliary gases for all the operating
processes, and receives the carrier gas from the GC unit used to transfer the sample to the
injector.
• The
auxiliary gases
used with the instrument are helium and nitrogen.
• The
carrier gas
used with the GC are helium, nitrogen, hydrogen, air, argon,
argon/methane. Other gases are rarely used.
Computer *
1365
400
Monitor *
85
25
* Power requirements vary by manufacturer. ** Single power module *** Two power modules
Table 5.
Maximum Heat Generated by Each Instrument (Sheet 2 of 2)
Equipment
Heat Output (BTU per Hr.)
Heat Output (in W)
Содержание TriPlus 300
Страница 4: ......
Страница 12: ...Contents xii TriPlus 300 Headspace Preinstallation Requirements Guide Thermo Scientific ...