Table 3-2.
LVSL available liners
Application
P/N
Glass liner packed with deactivated glass wool
453 220 65
Deactivated laminar liner (optional)
453 220 67
Chapter 3
Large Volume Splitless Injector Maintenance
Maintaining a Large Volume Splitless Injector
Maintenance and Troubleshooting Manual
57
If the glass liner breaks inside the injector
When replacing or removing a glass liner, it might break inside the injector. In this
case the broken parts of the liner must be removed from the injector, including the
glass splinters that might fall into the lower part of the vaporization chamber.
Refer to
Operating Sequence on page 67 for instructions.
When replacing the split line tubing
Replace the split line tubing when strong contamination or block has been
verifyied. The Kit Assy Split line (PN 190 507 10) is required.