Chapter 12
Maintaining an Electron Capture Detector
Clean the Collecting Electrode
226
Maintenance and Troubleshooting Manual
OPERATING SEQUENCE
Clean the Collecting Electrode
Materials needed
•
Screwdriver
•
2.5 mm Allen wrench
•
Tweezers
•
Ultrasonic cleaning bath
•
GC-grade hexane
•
GC-grade toluene
•
Fine emery paper
WARNING! Any maintenance or repair operations involving the radioactive source must be
performed ONLY by qualified personnel duly authorized to handle radioactive
material. Please contact your local Thermo Fisher Scientific Technical Service office.
WARNING! When handling organics solvents you must take precautions to avoid health
hazards. Refer to Appendix A,
for more information.
WARNING! This operation must be carried out at low temperature to avoid burns. Therefore,
before beginning the sequence, the detector must be cooled to room temperature.
PRECAUTIONS