Chapter 8
Maintaining a Purged Packed Column Injector
Clean or Replace the Liner
136
Maintenance and Troubleshooting Manual
OPERATING SEQUENCE
Clean or Replace the Liner
Materials needed
•
Glass liner (see item 8, Table 5-1)
•
Ultrasonic bath
•
Methanol/acetone mixture (1:1)
•
Tweezers
WARNING! This operation must be carried out at low temperature to avoid burns. Therefore,
before beginning the sequence, the injector must be cooled to room temperature.
PRECAUTIONS
WARNING! When handling organics solvents you must take precautions to avoid health
hazards. Refer to Appendix A,
for more information.
1. While in stand-by condition, press
OVEN
to
access the oven control table.
.
OVEN
Temp
80
80 <
Initial time
200
Ramp 1
Off
2. Scroll to
Temp
and press
OFF
.