Installing Optional Kits
Installing the Hot Injection Adapter Kit on the SSL/SSLBKF Injector
408
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
Figure 419.
Hot Injection Adapter (5)
19. Reconnect the column end to the injector and verify the connection point.
20. Open the gas supplies.
21. Power on the GC.
a.
Plug the power cable to the AC Input connector into the back of the GC and to the
wall outlet.
b. Flip out the power switch (breaker) to the position I.
22. Pressurize the module with the carrier gas.
23. Check for leaks.
a.
Use a handheld electronic leak detector (Thermo Scientific GLD Pro leak detector or
equivalent) to check the two fitting for leaks.
b. If you detect a leak, tighten the connection and retest it.
c.
Repeat this process until all connections are leak free.
24. Close the front door of the GC.
25. If present, update the autosampler for the new injection position.
Note
If only a port of the adapter is connected to the device for the sampling of gas,
close the other one with the blind nut provided.
IN
OUT