4
Performing Detectors Routine Maintenance
Maintaining a Flame Photometric Detector (FPD)
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
185
Figure 170.
Interferential Filter Removal
8. Clean the filter.
a.
Using a clean paper towel, clean the filter on both faces.
9. Reassembly the filter and the photomultiplier assembly.
a.
Insert the cleaned filter or a new filter into its housing. The mirror face must be
oriented towards the flame. See
CAUTION
Filters are fragile. Pay attention not to let the filter fall down and damage.
CAUTION
If the filter installed is equipped with the spacer (see the kits PN 19050785 and
PN 19050786), first remove the spacer, and then the interferential filter.
Keep them using a clean paper towel.
Spacer
Interferential Filter
CAUTION
Avoid touching the filter with your fingers. If you see fingertips on the filter,
clean it using GC-grade methanol and air dry before remounting.
Содержание TRACE 1300
Страница 1: ...TRACE 1300 and TRACE 1310 Gas Chromatographs Hardware Manual PN 31715002 Revision F January 2016 ...
Страница 4: ......
Страница 16: ......
Страница 94: ......
Страница 224: ......
Страница 450: ......
Страница 578: ......
Страница 610: ......
Страница 620: ......
Страница 624: ......