Performing Detectors Routine Maintenance
Maintaining a Flame Photometric Detector (FPD)
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
183
a.
Plug the power cable to the AC Input connector into the back of the GC and to the
wall outlet.
b. Flip up the power switch (breaker) to the position I.
11. Set the normal working conditions.
Cleaning or Replacing the FPD Interferential Filter
Before maintaining the detector, read the following precautions:
To clean or replace the FPD interferential filter
1. Put the GC in standby condition.
2. Switch off the flame. Hydrogen and air are automatically closed.
3. Cool the detector to room temperature.
4. Power off the GC.
a.
Push down the power switch (breaker), located at the back of the instrument, to the
position O.
b. Unplug the power cable from the AC Input connector into the back of the GC and
from the wall outlet.
5. If other external modules are present, unplug the power cable from the AC Input
connector of each module, and from the wall outlet.
6. Disconnect the signal, excitation voltage, and ignition/heating cables from the detector.
See
WARNING
Carry out all the operations at low temperature to avoid burns. Therefore,
before beginning maintenance, cool the detector to room temperature.
CAUTION
When handling organic solvents you must take precautions to avoid health
hazards.
Note
By pressing the
Maintenance
button, the GC cool down is automatically
carried out.